Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8021B: (2 year) CAL VER /5500 INSTRUMENT: Fluke 8022B: (2 year) CAL VER /5500 DATE: 10-Feb-95 AUTHOR: Fluke Corporation REVISION: $Revision: 1.2 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 29 NUMBER OF LINES: 120 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Fluke 8021B instruction manual (PN 616037 July 1981). # Fluke 8022B instruction manual (PN 616045 July 1981). # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 55cw8021.bmp # 55_v8021.bmp # 55_a8021.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5500A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F W 1.002 ASK+ K 1.003 HEAD EQUIPMENT SETUP 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP Ambient temperature: 21C - 25C. 1.005 DISP Relative humidity: less than 80%. 1.005 DISP Allow the UUT to stabilize. 1.006 DISP Turn the UUT on. 1.007 HEAD {RESISTANCE TESTS} 1.008 PIC 55CW8021 1.009 HEAD RESISTANCE TESTS: {2 kOhm Range} 1.010 5500 2 0.000kZ +0.001U CW 2.001 HEAD RESISTANCE TESTS: {200 Ohm Range} 2.002 5500 200 0.0Z +0.2U CW 3.001 5500 200 190.0Z 0.9U CW 4.001 HEAD RESISTANCE TESTS: {2 kOhm Range} 4.002 5500 2 1.900kZ 0.005U CW 5.001 HEAD RESISTANCE TESTS: {20 kOhm Range} 5.002 5500 20 19.00kZ 0.05U CW 6.001 HEAD RESISTANCE TESTS: {200 kOhm Range} 6.002 PIC 55_V8021 6.003 5500 200 190.0kZ 0.5U 2W 7.001 HEAD RESISTANCE TESTS: {2000 kOhm Range} 7.002 5500 2000 1900kZ 39U 2W 8.001 HEAD RESISTANCE TESTS: {20 MOhm Range} 8.002 5500 20 10.00MZ 0.2U 2W 9.001 JMPT 11.003 Fluke 8022B: (2 year) CAL VER /5500 9.002 HEAD {CONTINUITY TEST AUDIBLE TONE} 9.003 RNG 2kZ 9.004 DISP Depress the AC/DC switch to activate the audible tone. 9.005 5500 0Z S 2W 9.006 EVAL Is there an {audible tone}? 10.001 DISP Disconnect the 5500A from the UUT. 10.002 DISP Apply line voltage (110V @ 60Hz) to the V Ohm S 10.002 DISP and COMMON inputs using an appropriate cable. 10.003 EVAL Is UUT {beeping at a rate of approximately 5 to 10 per sec}. 11.001 DISP Disconnect the line voltage from the UUT. 11.002 PIC 55_V8021 11.003 HEAD {DC VOLTAGE TESTS} 11.004 HEAD DC VOLTAGE TESTS: {200mV Range} 11.005 5500 200 190.0mV 0.6U 2W 12.001 5500 200 -190.0mV 0.6U 2W 13.001 HEAD DC VOLTAGE TESTS: {2V Range} 13.002 5500 2 0.000V 0.001U 2W 14.001 5500 2 1.900V 0.006U 2W 15.001 HEAD DC VOLTAGE TESTS: {20V Range} 15.002 5500 20 19.00V 0.06U 2W 16.001 HEAD DC VOLTAGE TESTS: {200V Range} 16.002 5500 200 190.0V 0.6U 2W 17.001 HEAD DC VOLTAGE TESTS: {1000V Range} 17.002 5500 1000 1000V 3U 2W 18.001 HEAD {AC VOLTAGE TESTS} 18.002 HEAD AC VOLTAGE TESTS: {200mV Range} 18.003 RNG 200mV AC 18.004 5500 0.0mV S 2W 18.005 MEMI Enter UUT reading in millivolts AC: 18.006 MEME 18.007 MEMC mV +0.2U 19.001 5500 200 190.0mV 1.6U 100H SI 2W 20.001 5500 200 19.0mV 0.3U 100H SI 2W 21.001 HEAD AC VOLTAGE TESTS: {2V Range} 21.002 5500 2 1.900V 0.016U 100H SI 2W 22.001 HEAD AC VOLTAGE TESTS: {20V Range} 22.002 5500 20 19.00V 0.16U 100H SI 2W 23.001 HEAD AC VOLTAGE TESTS: {200V Range} 23.002 5500 200 190.0V 1.6U 100H SI 2W 24.001 HEAD AC VOLTAGE TESTS: {750V Range} 24.002 5500 750 750V 9U 400H SI 2W 25.001 HEAD {DC CURRENT TESTS} 25.002 PIC 55_A8021 25.003 HEAD DC CURRENT TESTS: {2mA Range} 25.004 5500 2 1.900mA 0.015U 2W 26.001 HEAD DC CURRENT TESTS: {20mA Range} 26.002 5500 20 -19.00mA 0.15U 2W 27.001 HEAD DC CURRENT TESTS: {200mA Range} 27.002 5500 200 190.0mA 1.5U 2W 28.001 HEAD DC CURRENT TESTS: {2000mA Range} 28.002 5500 2000 1900mA 15U 2W 29.001 END